摘要 | A method for the convenient detection of lead at the parts-per-billion (ppb)-level has been developed; it uses a conventional compact disc (CD) as the platform for preparing DNAzyme assays and an unmodified optical drive of ordinary desktop/laptop computers as the readout device. In particular, by immobilization of Pb(2+) -specific DNAzyme sensing constructs on the "transparent side" of a conventional CD-R via mild surface reactions, the Pb(2+) concentration can be determined by a free diagnostic program that checks the error distribution on the CD (i.e., it extracts the number of errors in a prerecorded audio file). The reading errors increase monotonically over a wide range of Pb(2+) concentrations (from 10 nM to 1 mM), and the selectivity is confirmed by testing several other divalent cations (Zn(2+), Ba(2+), Mg(2+), Ca(2+), Cu(2+), and Hg(2+)),; A method for the convenient detection of lead at the parts-per-billion (ppb)-level has been developed; it uses a conventional compact disc (CD) as the platform for preparing DNAzyme assays and an unmodified optical drive of ordinary desktop/laptop computers as the readout device. In particular, by immobilization of Pb2+ -specific DNAzyme sensing constructs on the "transparent side" of a conventional CD-R via mild surface reactions, the Pb2+ concentration can be determined by a free diagnostic program that checks the error distribution on the CD (i.e., it extracts the number of errors in a prerecorded audio file). The reading errors increase monotonically over a wide range of Pb2+ concentrations (from 10 nM to 1 mM), and the selectivity is confirmed by testing several other divalent cations (Zn2+, Ba2+, Mg2+, Ca2+, Cu2+, and Hg2+), |
WOS标题词 | Science & Technology ; Physical Sciences |
关键词[WOS] | ATOMIC-ABSORPTION-SPECTROMETRY ; IN-VITRO SELECTION ; RNA-CLEAVING DEOXYRIBOZYME ; CATALYTIC BEACON SENSOR ; GOLD NANOPARTICLES ; WATER SAMPLES ; METAL-IONS ; DNA ; FLUORESCENCE ; SENSITIVITY |
收录类别 | SCI |
语种 | 英语 |
WOS研究方向 | Chemistry |
WOS类目 | Chemistry, Analytical |
WOS记录号 | WOS:000287685800012 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://210.75.249.4/handle/363003/1582 |
专题 | 中国科学院西北高原生物研究所 |
作者单位 | 1.Simon Fraser Univ, Dept Chem, Burnaby, BC V5A 1S6, Canada 2.Chinese Acad Sci, NW Inst Plateau Biol, Xining 810001, Qinghai, Peoples R China |
推荐引用方式 GB/T 7714 | Wang, Honglun,Ou, Lily M. L.,Suo, Yourui,et al. Computer-Readable DNAzyme Assay on Disc for ppb-Level Lead Detection[J]. ANALYTICAL CHEMISTRY,2011,83(5):1557-1563. |
APA | Wang, Honglun,Ou, Lily M. L.,Suo, Yourui,&Yu, Hua-Zhong.(2011).Computer-Readable DNAzyme Assay on Disc for ppb-Level Lead Detection.ANALYTICAL CHEMISTRY,83(5),1557-1563. |
MLA | Wang, Honglun,et al."Computer-Readable DNAzyme Assay on Disc for ppb-Level Lead Detection".ANALYTICAL CHEMISTRY 83.5(2011):1557-1563. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Computer-Readable DN(3227KB) | 开放获取 | -- | 浏览 下载 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论